With the Fluke 725 Multifunction Process Calibrator you can easily source and measure almost any process parameter and test almost any process device
Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.
From the moment you pick it up, the Fluke 725 is ready to perform as a powerful, yet easy-to-use field calibrator. Its simple controls, without menus, make operation easy.
Other useful features:
- Powers transmitters during test using loop supply with simultaneous mA measurement
- Handles fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
- Provides a backlit display to provide better visibility in poor light
- Features small, streamlined shape, and rugged, reliable design that stands up to field use
- Sources mA with simultaneous pressure measurement to conduct valve and I/P tests
- Supports flow meter testing with frequency and CPM functions
- Performs fast linearity tests with auto-step and auto-ramp features
Measurement Accuracy: |
||
Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
30.000 V | 0.02% + 2 counts (lower display) | |
100.00 mV | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA | 0.02% + 2 counts |
Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
Frequency | 2.0 to 1000.0 CPM | 0.05% + 1 count |
1.0 to 1100.0 Hz | 0.05% + 1 count | |
1.00 to 10.00 kHz | 0.05% + 1 count | |
Sensitivity | 1 V peak-to-peak-minimum | |
Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |
Source Accuracy: |
||
Voltage DC | 100.00 mV | 0.02% + 2 counts |
10.000 V | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA (source) | 0.02% + 2 counts |
24.000 mA (simulate) | 0.02% + 2 counts | |
Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
Frequency | 2.0 to 1000.0 CPM | 0.05% |
1.0 to 1100.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.25% | |
Waveform | 5 V p-p squarewave, -0.1 V offset | |
RTDs and Thermocouples: |
||
Measure accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.8°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
XK | 0.6°C | |
BP | 1.2°C | |
Source accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
Accuracy stated for 4-wire measurement. | ||
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.4°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
XK | 0.6°C | |
BP | 1.2°C | |
Specifications: |
||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
Loop power function | Voltage | 24 V |
Accuracy | 10% | |
Maximum current | 22 mA, short circuit protected | |
Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
Steps | 25% of range, 100% of range | |
Environmental Specifications: |
||
Operating temperature | -10°C to +55°C | |
Storage temperature | -20°C to 71°C | |
Operating altitude | 3000 m | |
Safety Specifications | ||
Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
Mechanical and General Specifications: |
||
Size | 130 x 236 x 61 mm | |
Weight | 0.65 kg | |
Batteries | 4 AA alkaline batteries | |
Battery replacement | Separate battery compartment, accessable without breaking calibration seal | |
Side port connections | Pressure module connector, also used for remote real-time programming |
Documentation:
Reviews