Hart Scientific 9103: Dry Block Calibrator

$0.00

The Hart Scientific 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C, and its display is calibrated to an accuracy of ±0.25 °C at all temperatures within its range. The 9103 reaches temperatures 50 °C below ambient, so –25 °C is reached under normal ambient conditions.

SKU: hart-scientific-9103 Categories: , ,

More Information

The Hart Scientific 9103 Field Dry Block Calibrator is portable, flexible, and suitable for high-volume calibrations or certifications

The 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C, and its display is calibrated to an accuracy of ±0.25 °C at all temperatures within its range. In just eight minutes, 0 °C is reached, and 100 °C is reached in six minutes, so your time is spent calibrating—not waiting. The 9103 reaches temperatures 50 °C below ambient, so –25 °C is reached under normal ambient conditions.

Choose one of three removable inserts sized for probes from 1/16 inch to 1/2 inch in diameter. Insert A handles a full range of probe sizes with a single well of each size. Insert B features two wells each of 3/8, 1/4, and 3/16 inches in diameter for doing comparison calibrations. Insert C has six 1/4-inch-diameter wells for multiple probe calibrations, and Insert D has three pairs of metric sized wells.

Specifications

  9103  9140 
Range  –25 °C to 140 °C
(–13 °F to 284 °F)
at 23 °C ambient 
35 °C to 350 °C
(95 °F to 662 °F) 
Accuracy  ± 0.25 °C (holes greater than 1/4″ [6.35 mm]: ± 1 °C)  ± 0.5 °C (holes greater than 1/4″ [6.35 mm]: ± 1 °C) 
Stability  ± 0.02 °C at –25 °C
± 0.04 °C at 140 °C 
± 0.03 °C at 50 °C
± 0.05 °C at 350 °C 
Well-to-Well Uniformity  ± 0.1 °C between similarly sized wells  ± 0.1 °C with similarly sized wells 
Heating Times  18 minutes from ambient to 140 °C  12 minutes from ambient to 350 °C 
Cooling Times  20 minutes from ambient to –25 °C  15 minutes from 350 °C to 100 °C 
Stabilization Time  7 minutes 
Immersion Depth  124 mm (4.875″) 
Inserts  Insert A, B, C, or D included (specify when ordering) 
Outside Insert Dimensions  31.8 mm dia. x 124 mm (1.25 x 4.88 in) 
Computer Interface  RS-232 interface
Power  115 V ac (± 10 %), 1.3 A or 230 V ac (± 10 %), 0.7 A, switchable, 50/60 Hz, 150 W  115 V ac (± 10 %), 4.4 A or 230 V ac (± 10 %), 2.2 A, switchable, 50/60 Hz, 500 W 
Size (WxHxD)  143 x 261 x 245 mm
(5.63 x 10.25 x 9.63 in) 
152 x 86 x 197 mm
(6 x 3.375 x 7.75 in) 
Weight  5.7 kg (12 lb)  2.7 kg (6 lb) 
NIST-Traceable Certificate  Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, and 140 °  Data at 50 °C, 100 °C, 150 °C, 200 °C, 250 °C, 300 °C, and 350 °C 