Fluke 718 300G Key features:

  • Pressure Source and Milliamp Measurement to calibrate and maintain almost any pressure device
  • Integrated pump is easily cleaned when accidently exposed to fluids that reduces cost of ownership and repairs and enables servicing the pump in the field
  • Error % calculation for quicker pass/fail decision making in the field
  • Best in class mA accuracy of 0.015% for better measurement confidence and workload coverage
  • Min/Max/Hold to capture changing measurements
  • Switch test feature for quicker and more accurate pressure switch testing
SKU: fluke-718-300g Categories: ,

More Information

The Fluke 718 300G Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges and switches up to 300 psi

This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.

Technical Specifications:

Fluke 718 Series:
Measure pressure (Internal sensor)
Accuracy 0.025% full scale
Functions Zero, min/max, hold, damp
Dry air, non-corrosives only
Fluke 718 300G:
Range -12 psi to 300 psi, (-850 mbar to 20 bar, -85 to 2068.42 kPa)
Resolution 0.01 psi, 1 mbar, 0.1 kPa
Over pressure 375 psi 25 bar
Functions Zero, Min, Max, Hold, Damp
Fluke 718 series via 700 Series Pressure Modules:
Range 29 pressure modules, 0 – 1 inH20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs
Resolution Per pressure module specs
Accuracy To 0.025% of full span, per pressure module specs
Functions Zero, Min, Max, Hold, Damp
Media compatibility per pressure module specs:
Fluke 718 Series Via Built in Pump
Range -12 psi or -850 mbar to full scale Supported Pressure Units; psi, inH20 (4°C), inH20 (20°C), cmH20 (4°C), cmH20 (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop power range 24 V DC
Accuracy ±10%
Drive; 20 mA into 1000 Ω for battery > 6.8 V; 700 Ω for battery 5.8 to 6.8 V:
mA Measure:
Range 0 mA to 24 mA
Resolution 0.001 mA
Accuracy 0.015% + 1 count
General Specifications:
Maximum voltage 30 volts (non-operating)
Temperature -40°C to 60°C
Operating temperature -10°C to 55°C
Relative humidity 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude 3,000 M max
Shock 1 M drop test
Vibration Random, 2 g, 5 Hz – 500 Hz
Safety CSA C22.2 No. 1010.1: 1992
EMC EN50082-1: 1992 and EN55022: 1994 Class B
Size 216 x 94 x 66 mm (8.50 x 3.72 x 2.60 in)
Weight 992 g (35 oz)
Power, battery life 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Battery life 4 to 20 hours typical, depending on functions used
Warranty Three years (one year on pressure pump in Fluke 718)
Display LCD, 5 digit pressure and current simultaneous
For use with non-corrosive gasses